bASUnt“ is displayed. This is where you can set the required
weighing unit.
Press to change between the individual units and
press to select the required unit (ex. here: „kg“).
(example)
(example)
SCL.CAP 1”will appear briefly.
Setting first range
A numeric value appears and the digit on the right is
flashing.
Press to increase the numeric value or to
decrease. Press to change to a different decimal
place.
Set weighing scale capacity (here is an example of 50 kg)
and press to confirm.
(example)
rESOLE 1”will appear briefly followed by a numeric value.
Set readability for first range. You can select between
default values that are dependent on the max load of the
weighing scale.
Press to change between values.
Press to confirm the required value.
rESOLE 2”will appear briefly followed by a numeric value. As
above, enter required readability and press to confirm
the required value.
F” will appear briefly before the weighing scale changes to
adjusting mode and “CAL” will be displayed.
Linearizing a weighing scale (see chap 6.7.1)