
Ziegelei 1
D-72336 Balingen
Tel: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www. kern-sohn.com
FH_SO-BA-e-1217 2
3. Electrical Power Supply
Either by rechargeable battery or current power supply
Current power supply:
- Connection by power adapter
- Rechargeable batteries are charged
simultaneously
Rechargeable battery pack for mobile applications:
- Type: Ni 8.4V / 600 mAh
- Charging time: approx 1 hour
4. Technical Data
- Accuracy: ± 0,5 % of Capacity
- Data Sampling Rate: 2.000 Hz
- Weight: 640 g
5. Operation
Display
(1) Measuring Result
(2) Measuring Units
(3) Activation of PRINT Function
(4) Indication of power charging status
PEAK or AUTO-PEAK Mode
(5) Average value of stored peak values
(6) Force direction
(7) Occupancy of storing spaces
(8) AVERAGE- or Saving Mode
Operating keys
ON / OFF:
ON / OFF key
(For ON, press 1 sec.)
UNIT:
- Press short: Select unit:
N, kg or lb
- Press for 2 sec.: Display return
ZERO:
Three functions:
- Zeros the measuring result (Tara function)
- Cleans the peak value (in Peak mode)
- Saves a setting (in SET mode
SET:
- 1 x Press: Upper Limit [HidT / Hi Lt]. To
change press: ▲ or ▼ (see section
High/Low Limit function)
- 1 x Press: Lower Limit [LodT]. To change
press ▲ or ▼ (see section High/Low Limit
function )
- 1 x Press: Minimum limit to save peak
values in the instrument’s memory.
[LE.SET / Lo Pe]. Please see section „Min
Limit Peak Save“. To change press: ▲ or
▼ (Only in „Peak-Mode“ active) (see
section min. limit peak save)
- 1 x Press: Sets the Auto-STOP force value
that stops the movement of a test stand.
Adjustment of the force value. To change
press: ▲ or ▼ (only active in „Battery-
Mode“)
- 1 x Press: Sets the AUTO-OFF force value
that stops the movement of a test stand. To
change press: ▲ or ▼ (Only in „Battery-
Mode“ active)
- 1 x Press: Peak-Freeze-Time [PE.2E /
A.PE / HoldT]: Time period in which a peak
value is being shown in the display in sec.
To change press: ▲ or ▼
Comentarios a estos manuales