Kern-sohn SMT 10K-3 Manual de usuario Pagina 1

Busca en linea o descarga Manual de usuario para Escamas Kern-sohn SMT 10K-3. KERN & SOHN SMT 10K-3 User Manual [it] [es] [en] [de] Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 37
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
email: info@kern-sohn.com
Phone: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Operating instructions
Bottleweighing Balance
KERN SMT
Version 1.4
12/2014
GB
SMT-BA-e-1414
Vista de pagina 0
1 2 3 4 5 6 ... 36 37

Indice de contenidos

Pagina 1 - Bottleweighing Balance

KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen email: [email protected] Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-

Pagina 2 - KERN SMT

SMT-BA-e-1414 10

Pagina 3 - 1 Technical Data

6.2 Select operator and location Select operator: Press the operator button and an option display for admitted operators will appear. Confirm th

Pagina 4

Select location: Press the Location button and the option displays of the available locations will appear. Confirm the correct location by pressin

Pagina 5 - 3 Basic Safety Precautions

Press the start button to trigger this screen asking you to select a drink. Drink may be selected via SRCH-CODE SRCH-NAME AUTO-LIST or SCAN funct

Pagina 6

Enter the name of the drink. In the second information line the information will appear if a selection via selecting fields will be available. Plea

Pagina 7

Put the selected drink on the balance. Now the contents of the bottle will be displayed as a result in cl. Press the “FRAC/VOL” button and the p

Pagina 8 - 6 Operation

After pressing the "CLR / BACK" button, the last inventory will be displayed. To delete the last inventory "Clear Registry" wil

Pagina 9 - 9 SMT-BA-e-1414

Now in the second information line the first listed drink appears. Place the specified bottle. The bottle content will be displayed in „CL“.

Pagina 10 - SMT-BA-e-1414 10

When the bottle is removed, the following drink appears. Use the „BACK/DRNK“ button to register an other bottle of the previous drink. Use the „

Pagina 11

Now select the suitable scanner. When the desired scanner is selected, confirm via the return button. Table Scanner Datalogic 19 SMT-BA-e-1414

Pagina 12

GB KERN SMT Version 1.4 12/2014 Operating instructions Bottleweighing Balance Table of Contents 1 Technical Data ...

Pagina 13

Handheld Scanner Datalogic Quick Scan Mobile SMT-BA-e-1414 20

Pagina 14

6.8 Activate automatic storage results: To determine the automatic storage of results from the basic screen, press the „MENU“ button. If the a

Pagina 15

Now press the „Basic settings“ button. Press the „Store Result:Auto“ button and acknowledge with Enter. 6.9 Input of operator and locations:

Pagina 16

Now press the „Configuration“ button. Now press the „Operator“ button. To enter a location press the „Location“ button. With the help of the

Pagina 17

To enter a new operator, overwrite no longer required operators or call up empty description fields by pressing them. Enter the name via the inp

Pagina 18

Enter the location via the input keyboard field. Save the location via the Enter button. 6.10 Enabling / Disabling the zero passing: Press th

Pagina 19

Now the selection may take location via the zero-passing button. After that save with the enter button. 6.11 Activation of the automatic switc

Pagina 20

Now press the „Basic settings“ button. Now press the arrow button Set AUTO INPUT to "ON" and save. 27 SMT-BA-e-1414

Pagina 21

6.12 Data Transfer PC-SMT, SMT-PC SETUP Setup Balance Bring back balance into the basic screen (STOP/TAKE). Connect the PC to the bala

Pagina 22

Setup Application Requirements: • Windows XP or higher • Microsoft Office 2003 or higher • .Net Framework 2.0 or higher 1. Go to Settings Menu

Pagina 23

1 Technical Data KERN SMT 10K-3 Readability (d) 1 g Weighing range (max) 10 000 g Taring range (subtractive) 10 000 g Reproducibility 1 g Line

Pagina 24

UPLOAD DATA 1. Make sure balance is in basic Mode 2. Choose your File with File → Open SMT-BA-e-1414 30

Pagina 25

3. Press „Upload from PC > balance“ Button to Upload the chosen File Information: It will upload the same file everytime, till you select ano

Pagina 26

DOWNLOAD DATA 1. Make sure balance is in basic Mode 2. Press „Download balance PC“ Button 3. You'll be asked for a filename SMT-BA-e-14

Pagina 27

4. Excel will automatically open containing the transfered Data 33 SMT-BA-e-1414

Pagina 28

7 Data output RS 232 C Technical data: 8-bit ASCII Code  1 start bit, 8 data bits, 1 stop bit, no parity bit  Baud rate selectable from 2400,

Pagina 29

8 Service, maintenance, disposal 8.1 Cleaning Before cleaning, please disconnect the appliance from the operating voltage. Please do not use agg

Pagina 30

9 Instant help In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process mus

Pagina 31

10 Declaration of conformity KERN & Sohn GmbH D-72322 Balingen-Frommern Postfach 4052 E-Mail: [email protected] Tel: 0049-[0]7433- 9933-0 Fax:

Pagina 32

2 Basic Information (General) It is absolutely necessary that you read and understand the operating instructions prior to installation and commissi

Pagina 33

3 Basic Safety Precautions 3.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and co

Pagina 34 - 1

5 Unpacking, Setup and Commissioning 5.1 Installation Site, Location of Use The balances are designed in a way that reliable weighing results are a

Pagina 35

5.6 Adjustment As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance

Pagina 36 - 9 Instant help

6 Operation 6.1 Starting balance by touch screen Welcome display: After touching the touch screen and starting the balance, this screen will app

Comentarios a estos manuales