Kern-sohn CCS Version 2.0 Manual de usuario Pagina 33

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 81
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 32
CFS/CCS-BA-e-1220 33
„LoAd 3“ will be displayed
50 kg adjustment weight placed carefully in the
centre of the weighing plate, wait for stability display
and confirm with .
After the adjustment the balance will carry out a self-
test. Remove adjusting weight during selftest,
balance will return into weighing mode automatically.
In case of fault or a wrong weight an error message
( ) will appear, then repeat the
linearization procedure.
8.2.1 Table linearisation points CFS 50K-3 models.
Max: 50 kg
Display
Weight
„L0Ad 0“
0 kg
„L0Ad 1“
15 kg
„L0Ad 2“
30 kg
„L0Ad 3“
50 kg
Vista de pagina 32
1 2 ... 28 29 30 31 32 33 34 35 36 37 38 ... 80 81