Procedure:
Ö Call-up menu setting , see chap. 1.
Ö Acknowledge with .
Ensure that there are no objects on the weighing plate.
Ø
Ö Wait for stability display „STABLE“, then press . When
“LoAd 1“ is displayed, put the first adjustment weight (1/3
max) carefully in the centre of the weighing platform.
Ö Wait for stability display „STABLE“, then press . When
“LoAd 2“ is displayed, put the second adjustment weight (2/3
max) carefully in the centre of the weighing platform.
Ö Wait for stability display „STABLE“, then press . When
“LoAd 3“ is displayed, put the third adjustment weight (1/3
max) carefully in the centre of the weighing platform.
Ö Wait for stability display „STABLE“, then press .
Ö After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, the appliance will
return into weighing mode automatically.
An adjusting error or incorrect adjusting weight will be indi-
cated by the error message; repeat adjustment procedure.