Ensure that there are no objects on the weighing
plate. Press the PRINT key.
In the display there appears the value of the
adjustment weight.
Place the required adjustment weight in the centre of
the sample dish, then press the PRINT key. If the
adjustment was successful, this will be followed by
automatic reset to the menu.
Take away adjustment weight.
The ESC key may be used to cancel the adjustment.
Errors during adjustment or the use of an incorrect
adjusting weight will result in an error message.
Repeat adjustment.
Return to weighing mode:
Press the ESC-key repeatedly until the query
SAVE?“ appears.
Confirm query by pressing the PRINT-key or reject it
by pressing the ESC-key.
Setup
!
Saveing...
Setup29.01.08
We recommend:
Carry out all adjustments as close as possible to the maximum load of the balance.