MLS_C-BA-e-1330 49
To return to menu, press the PRINT-key. This is where
you can store further drying parameters.
or
Return to weighing mode:
Press the ESC-key repeatedly until the query ”SAVE?“
appears. Confirm query by pressing the PRINT-key or
reject it by pressing the ESC-key.
2. Shutoff criterion
This is used to select the criteria used for finishing the drying process.
Press the button; the menu item “Finish“ appears.
Press -button
Press the arrow keys ( ), to select among the
finishing criteria:
Shutoff criterion mg per time
Drying is shut off when the weight loss for the set
time is smaller than the number of set digits (1 digit
= 1mg)
How to select the desired setting:
Press -button
Press the arrow keys ( ), to carry out the desired
setting.
Automatic 1: Shutoff for change of weighing value
< 1mg within 10s.
Automatic 2: shutoff during change of weighing value
< 1mg within 25s
Automatic 3: Shutoff for change of weighing value
< 1mg within 60s.
Automatic 4: shutoff during change of weighing value
< 1mg within 90s
Automatic 5: shutoff during change of weighing value
< 1mg within 120s
Comentarios a estos manuales